The operating parameters menu contains the High Limit parameter. The High
Limit parameter adjusts the upper set-point temperature. The factory default
and maximum are set to 126 °C. For safety, a user can adjust the High Limit
parameter down so the maximum temperature set-point is restricted.
Press “SET” to enable adjustment of the High Limit parameter.
HL Flashes “HL” and then displays the setting
H=126 Current HL setting
Adjust the HL parameter using “UP” or “DOWN”.
H=90 New High Limit setting
Press “SET” to accept the new High Limit parameter.
7.12 Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
SEriAL Serial RS-232 interface parameters menu
The serial interface parameters menu contains parameters which determine the
operation of the serial interface. These controls only apply to instruments fitted
with the serial interface. The parameters in the menu are: BAUD rate, sample
period, duplex mode, and linefeed.
Press “SET” to access the serial parameters.
7.12.1 BAUD Rate
The BAUD rate is the first parameter in the menu. The BAUD rate setting de
-
termines the serial communications transmission rate.
The BAUD rate parameter is indicated by,
bAUd
Flashes “bAUd and then displays the setting
2400 b Current BAUD rate
The BAUD rate of the serial communications may be programmed to 300 600,
1200, 2400, 4800, or 9600 BAUD. 2400 BAUD is the default setting. Use “UP”
or “DOWN” to change the BAUD rate value.
4800 b New BAUD rate