Loading ...
Loading ...
Loading ...
3 Specifications and Environmental
Conditions
3.1 Specifications
Operating Range –25°C to 140°C (–13°F to 284°F), at 23°C ambient
Accuracy ±0.25°C (±1.0°C in holes > ¼” [6.4 mm])
Stability ±0.02°C at –25°C
±0.04°C at 140°C
Well-to-Well Uniformity ±0.1 with similar probes
Resolution 0.1°C or °F
Display Scale °F or °C, switchable
Thermal Electric Devices (TED) 150 W
Heating Time 18 minutes from ambient to 140°C
Cooling Time 20 minutes from ambient to –25°C
Stabilization Time 7 minutes
Immersion Depth 4.875 inches (124 mm)
Power Requirements 115 VAC (±10%), 1.3 A or 230 VAC (±10%) 0.7 A, switchable, 50/60
Hz, 150W
Inserts A, B, or C
Computer Interface RS-232 included with free Interface
-it
Software (Model 9930)
Size 10.25” H x 5.63” W x 9.63” D
(26.1 x 14.3 x 24.5 cm)
Weight 12 lb. (5.7 kg)
3.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trou
-
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and clean
-
ing recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
•
temperature range: 5–45°C (41–113°F)
•
ambient relative humidity: 15 - 50%
•
pressure: 75kPa - 106kPa
•
mains voltage within ± 10% of nominal
•
vibrations in the calibration environment should be minimized
•
altitude less than 2,000 meters
9
3 Specifications and Environmental Conditions
Specifications
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...