SDG2000X Service Manual.doc
55
Vertical Accuracy Calibration
This adjustment aims to calibrate the amplitude at low frequency, using a DMM
with high precision. The SDG2000X begins this portion of the calibration after
the completion of the self adjustment. It calibrates the Sine wave with a 10 kHz
frequency and selects several amplitudes to calibrate. Compare the measured
amplitude and the standard to obtain a set of compensation values.
1. Connect the DMM, PC and SDG2000X generator as shown below:
2. Double click and run the script VertAccCalib.py prepared previously.
3. Change the channel connection according to the prompt message.