4.Advanced User Guidebook
The description of Pass/fail Menu is shown as the follows:
Pass/Fail test:
Detect whether the input signal is within the limits of the rule, if it exceeds limits of
the rule, it is "Fail"; otherwise it is "Pass". Also it can output fail or pass signal by
built-in and configurable output port. To run the test, read the following steps:
1. Push the Utility button, select Function in the bottom menu, select Pass/fail in the
left menu.
2. Enable switch on: Select Operate in the bottom menu, select Enable in the right
menu as ON.
3. Create rule: Select Rule in the bottom menu. Select Source in the right menu,
select the source in the left menu. Set Horizontal tolerance and Vertical tolerance
in the right menu. Select Create in the right menu to create the rule.
4. Set output type: Select Output in the bottom menu to enter output option setting.
Choose any one or two of the options "Pass", "Fail" or "Beep". "Pass" and "Fail"
are mutually exclusive options, which could not be chosen simultaneously. "Stop"
means stop once the condition satisfies your setting.
5. Begin to test: Select Operate in the bottom menu, select Operate in the right
menu as Start, the test will begin.
6. Save rule: Select SaveRule in the bottom menu. Select the save location in the
left menu, and then select Save in the right menu to save the rules, which could be
called up at once when need. Select Load to call up the rule saved.
Note:
1. When Pass/Fail is ON, if XY or FFT is ready to run, then Pass/Fail will be closed;
under the mode of XY or FFT, Pass/Fail is unable.
2. Under the mode of Factory, Auto Scale and Auto Set, Pass/Fail will be closed.
3. When no save setting left in the rule save, tip will be given to show "NO RULE
SAVED".
Function Menu
Setting
Description
operate
Enable
Operate
Control enable switch
Control operate switch
Output
Pass
Fail
Beep
Stop
Info
Signal tested corresponds with the rule
Signal tested not correspond with the rule
Beep when it satisfies the rule
Stop once satisfying the rule
Control the display status of info frame
Rule
Source
Horizontal
Vertical
Create
Select source CH1, CH2 or Math
Change the Horizontal tolerance value by turning the
M knob
Change the Vertical tolerance value by turning the M
knob
Use the rule set as testing rule
SaveRule
Number
Save
Load
Select any one from Rule1 - Rule8 as your rule name
Select Save to save the rule
Load some rule as the testing rule