5.Advanced User Guidebook
26
(3) Press the F1 button to select the Band Limit as full band. The high
frequency of the signal will be allowed to pass.
(4) Press the F2 button to select the Band Limit as 20M. The bandwidth is
limited to 20 MHz. The frequencies above 20 MHz will be rejected.
Use Mathematical Manipulation Function
The Mathematical Manipulation function is used to show the results of the addition,
multiplication, division and subtraction operations between Channel 1 and Channel 2,
and the FFT operation of Channel 1 or Channel 2.
Figure 5-7 Wave math menu
The corresponding FCL (Functional Capabilities List) of the Waveform Calculation:
Function Menu
Setting
Description
Dual Wfm
Math
Factor1
CH1
CH2
Select the signal source of the factor1
Sign + - * / Select the sign of mathematical manipulation
Factor 2
CH1
CH2
Select the signal source of the factor2
VERT POS
Turn the M knob to adjust the vertical
position of M waveform
VOLTS/DIV
Turn the M knob to adjust the
voltage
division of M waveform
FFT
Source
CH1
CH2
Select CH1 as FFT source.
Select CH2 as FFT source.
Window
Hamming
Select window for FFT.