General Specications
Voltage, Terminal V:
DC Rang
e
AC Rang
e
Resolution Noise Floor Accuracy
(%+counts)
600 V 600 V 275 µV 9 mV 0.5 + 20
600 V 430 V 140 µV 5 mV 0.5 + 20
455 V 320 V 95 µV 3 mV 0.5 + 20
365 V 255 V 70 µV 3 mV 0.5 + 20
260 V 185 V 50 µV 2 mV 0.5 + 20
200 V 140 V 35 µV 1090 µV 0.5 + 20
135 V 95 V 25 µV 730 µV 0.5 + 20
60 V 45 V 20 µV 610 µV 0.5 + 20
40 V 30 V 10 µV 305 µV 0.5 + 20
30 V 22 V 6.4 µV 205 µV 0.5 + 20
25 V 18 V 4.8 µV 155 µV 0.5 + 20
18 V 12 V 3.2 µV 105 µV 0.5 + 20
14 V 10 V 2.4 µV 80 µV 0.5 + 20
9 V 6 V 1.6 µV 55 µV 0.5 + 20
1000 mV 700 mV 300 nV 10 µV 0.5 + 20
670 mV 475 mV 150 nV 5 µV 0.5 + 20
500 mV 350 mV 100 nV 4 µV 0.5 + 20
400 mV 275 mV 75 nV 3 µV 0.5 + 20
280 mV 200 mV 50 nV 2 µV 0.5 + 20
220 mV 150 mV 40 nV 2 µV 0.5 + 20
150 mV 100 mV 25 nV 1 µV 0.5 + 20
Current, Terminal A:
DC Rang
e
AC Rang
e
Resolution Noise Floor Accuracy
(%+counts)
10 A 9 A 3.8 µA 120 µA 0.5 + 50
8.5 A 6 A 1.9 µA 60 µA 0.5 + 50
6 A 4.5 A 1.3 µA 40 µA 0.5 + 50
5 A 3.5 A 1 µA 30 µA 0.5 + 50
3.5 A 2.5 A 0.7 µA 20 µA 0.5 + 50
2.75 A 2 A 0.5 µA 15 µA 0.5 + 50
1.75 A 1.25 A 0.4 µA 10 µA 0.5 + 50
Resistance
Range Resolution Noise Floor Accuracy
(%+counts)
1.0 + 50
1.0 + 50
1.0 + 50
1.0 + 50
1.0 + 50
1.0 + 50
1.0 + 50
1.0 + 50
1.0 + 50
1.0 + 50
1.0 + 50
1.0 + 50