SAFETY INFORMATION
To assure the correct use of the product, basic safety measures should always be
followed including the warnings and cautions listed in this instruction manual.
OPERATING THE DEVICE
Contact your physician for specific information about your blood pressure.
Self-diagnosis and treatment using measured results may be dangerous. Follow the
instructions of your physician or licensed healthcare provider.
DO NOT adjust medication based on measurement results from this blood pressure
monitor. Take medication as prescribed by your physician. Only a physician is
qualified to diagnose and treat High Blood Pressure.
The monitor is not intended to be a diagnostic device.
If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Contact a physician immediately.
If you have a condition that may compromise circulation, you may get an inaccurate
reading with this device. Consult your physician before using this device.
Read all of the information in the instruction manual and any other literature in the
box before operating the unit.
This device is intended for use in measuring blood pressure and pulse rate in the
adult population. Do not use this device on infants or persons who cannot express
their intentions.
Operate the device only as intended. Do not use the device for any other purpose.
Dispose of the device, components and optional accessories according to applicable
local regulations. Unlawful disposal may cause environmental pollution.
Do not use a cellular phone near the device. It may result in an operational failure.
Use only Omron authorized parts and accessories. Parts and accessories not
approved for use with the device may damage the unit.
Use only 1.5 V “AAA” alkaline batteries with this device. Do not use other types of
batteries. This may damage the unit.
CARE AND MAINTENANCE
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
Do not submerge the device or any of the components in water.
Store the device and the components in a clean, safe location.
Changes or modification not approved by the manufacturer will void the user
warranty. Do not disassemble or attempt to repair the unit or components.
SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL
WARNING
Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury to the user or patient or
damage to the equipment or other property.
BEFORE TAKING A MEASUREMENT
To ensure a reliable reading follow these recommendations:
1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes before
taking a measurement. Rest for at least 15 minutes before taking the measurement.
2. Stress raises blood pressure. Avoid taking measurements during stressful times.
3. The cuff can be applied to your left or right wrist.
4. Measurements should be taken in a quiet place.
5. Position the unit at heart level throughout the measurement.
6. Remain still and do not talk during the measurement.
7. Wait 2-3 minutes between measurements. The wait time allows the arteries to return to
the condition prior to taking the blood pressure measurement. You may need to
increase the wait time depending on your individual physiological characteristics.
8. Keep a record of your blood pressure and pulse readings for your physician. A single
measurement does not provide an accurate indication of your true blood pressure. You
need to take and record several readings over a period of time. Try to measure your
blood pressure at the same time each day for consistency.
INTRODUCTION
Thank you for purchasing the OMRON
®
BP652 IntelliSense
®
Automatic Wrist Blood
Pressure Monitor.
The monitor is compact and easy-to-use at home, at work, and during travel. Perfect for
people who frequently monitor their own blood pressure.
Your new digital blood pressure monitor uses the oscillometric method of blood pressure
measurement. This means the monitor detects your blood’s movement through your
brachial artery and converts the movements into a digital reading. An oscillometric
monitor does not need a stethoscope so the monitor is simple to use.
Clinical research has proven a direct relationship between blood pressure in the wrist and
blood pressure in the arm. Changes in wrist blood pressure reflect changes in arm blood
pressure because the arteries in the wrist and the arm are close to each other. Frequently
measuring the blood pressure in your wrist will provide your physician with an accurate
indication of changes in your blood pressure.
The BP652 comes with the following components:
•Monitor•StorageCase
•2“AAA”AlkalineBatteries
•InstructionManualprintedinEnglishandSpanish•QuickStartGuide
The BP652 IntelliSense
®
Automatic Wrist Blood Pressure Monitor is intended for home use.
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference. For specific information about your own blood pressure,
CONSULT YOUR DOCTOR.
Fill in for future reference.
DATE PURCHASED: __________________
SERIAL NUMBER: ____________________
SAVE THESE INSTRUCTIONS
INSTRUCTION MANUAL
Model
BP652
5329415-9B
7 series
Automatic Wrist Blood
Pressure Monitor
UNIT DISPLAY
DISPLAY SYMBOLS
IRREGULAR HEARTBEAT SYMBOL ( )
When the monitor detects an irregular rhythm two or more times during the
measurement, the Irregular Heartbeat Symbol ( ) will appear on the display with the
measurement values.
An irregular heartbeat rhythm is defined as a
rhythm that varies by less than 25% from the
average rhythm or 25% more than the average
rhythm detected while the monitor is measuring
the systolic and diastolic blood pressure.
If the irregular heartbeat symbol ( )
displays with your measurement results,
we recommend you consult your
physician. Follow the directions of your
doctor.
MOVEMENT ERROR SYMBOL ( )
TheMovementErrorSymbol()isdisplayedifyoumoveyourbodyduringthe
measurement. Please remove the wrist cuff, and wait 2-3 minutes. Reapply the wrist
cuff and take another measurement.
Internationally recognized guidelines for upper limit of
normal home blood pressure readings are as follows:
This is a general guideline because blood pressure depends on age and morbidity.
A lower home blood pressure goal may be recommended for certain patients,
including diabetic patients, pregnant women, and patients with renal failure.
WARNING
The monitor is not intended to be a diagnostic device.
WARNING
Contact your physician for specific information about your blood pressure.
Self-diagnosis and treatment using measured results may be dangerous.
Follow the instructions of your physician or licensed healthcare provider.
WARNING
DO NOT adjust medication based on measurement results from this blood
pressure monitor. Take medication as prescribed by your physician. Only a
physician is qualified to diagnose and treat High Blood Pressure.
Normal Heartbeat
Irregular Heartbeat
Short Long
Blood
pressure
Pulse
Pulse
Blood
pressure
Home Guideline for Upper Limit of Normal BP
Systolic Blood Pressure 135 mmHg
Diastolic Blood Pressure 85 mmHg
2
AAA
Alkaline Batteries
Instruction
Manual
KNOW YOUR UNIT
Main Unit:
Components:
Quick Start
Guide
SETTING THE DATE AND TIME
Set the monitor to the current date and time before taking a measurement for the first
time.
1. TO START THE SETTING
Press the Date/Time Setting button ( ) to adjust the date
and time.
The year flashes on the display.
2. SETTING THE YEAR
Press the Date/Time Setting button ( ) to advance by
increments of one year.
Press the Date/Time Setting button ( ) to set the current
year.
The month flashes on the display.
NOTES: The year can be set between 2009 and
2030. When the display reaches 2030, it
will return to 2009.
Press and hold the ( ) button to
increase the date and time values faster.
3. SETTING THE MONTH
Press the Date/Time Setting button ( ) to advance by
increments of one month.
Press the Date/Time Setting button ( ) to set the current
month.
The day flashes on the display.
4. SETTING THE DAY
Press the Date/Time Setting button ( ) to advance by
increments of one day.
Press the Date/Time Setting button ( ) to set the current
day.
The hour flashes on the display.
5. SETTING THE HOUR
The time is set using AM or PM.
Press the Date/Time Setting button ( ) to advance by
increments of one hour.
Press the Date/Time Setting button ( ) to set the current
hour.
The minute flashes on the display.
6. SETTING THE MINUTE
Press the Up button ( ) to advance by increments of one
minute.
Press the Date/Time Setting button ( ) to set the minute
setting.
7. TO TURN THE MONITOR OFF
Press the START/STOP button.
BATTERY REPLACEMENT
Low Battery Indicator
When the Low Battery Indicator appears on the display, turn the monitor off and
remove all the batteries. Replace with 2 new batteries at the same time. Long-life
alkaline batteries are recommended.
NOTES: Turn the unit off before replacing the batteries. If the batteries are removed
while the unit is still on, the date and time will be reset to that of the
previous use. The measurement results are not deleted.
Whenthebatteriesarereplaced,youmayneedtoresetthedateandtime.If
the year is flashing on the display, refer to Setting the Date and Time.
Thebatteriesincludedwiththedevicemayhaveashorterlife.
CAUTION
Dispose of the device, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental
pollution.
WARNING
If battery fluid should get in your eyes, immediately rinse with
plenty of clean water. Contact a physician immediately.
SETTING THE HEART GUIDE INDICATOR
Youcanchangetheoptionsforvarioussettingsonyourmonitor.PressingtheSET
buttontochangeasetting.PresstheMEMbuttontoselecttheoptionsforthatsetting.
After selecting a setting, press the START/STOP button to confirm the setting and turn
the power off.
SETTING THE HEART GUIDE INDICATOR
1. Select your User ID.
2.PresstheSETbutton.
NOTE: The default setting is on1.
3.PresstheMEMbuttontoselecttheHeartguideindicatoron1,on2,oroFF.
on1: After the Heart guide indicator lights in blue for more than 2 seconds, or in
orange for more than 5 seconds, measurement starts automatically, even
though the monitor is not in the proper position.
on2: The Heart guide indicator lights in blue for more than 2 seconds, measurement
starts automatically.
Measurement will not start if it lights in orange.
oFF: Measurement starts without the Heart guide indicator.
SETTING FOR RIGHT OR LEFT WRIST MEASUREMENT
1.Whileinthesettingmode,presstheSETbuttonuntil
the wrist symbol ( or ) appears on the display.
NOTES: The default setting is .
AfteryouchooseoFFfortheHeart
guide indicator, the monitor skips this
process.
2.PresstheMEMbuttontoselecttheleftwrist()ortherightwrist().
SETTING THE BUZZER
1.Whileinthesettingmode,presstheSETbuttonuntil
the buzzer symbol ( ) appears on the display.
NOTE: The default setting is on1 or on,
depending on the Heart guide indicator you
set.
2.PresstheMEMbuttontoselectthebuzzersetting.
If the Heart guide indicator is on1 or on2:
on1: The buzzer sounds only when the monitor is in proper position.
on2: The buzzer sounds a series of two short blips if your wrist is too far away
from the measuring position, and beeps when your wrist is in the proper
position.
oFF: The buzzer will not sound.
If the Heart guide indicator is oFF:
on: The buzzer sounds when pressing any buttons.
oFF: The buzzer will not sound.
RESETTING TO THE DEFAULT SETTING
The Heart guide indicator is set on1, the wrist for measurement is set , and the
buzzer is set on1 as default.
Toresettothedefaultsetting,pressandholdtheSETbuttonwhileinthesettingmode,
then press the START/STOP button simultaneously for more than 2 seconds.
TAKING A MEASUREMENT
5. Fold the remaining part of the wrist cuff back out of the
way.
APPLYING THE CUFF ON THE RIGHT WRIST
When taking a measurement using the right wrist position the
cuff as shown in the illustration.
Refer to Setting for Right or Left Wrist for Measurement.
TROUBLESHOOTING TIPS
PROBLEM
No power.
No display appears on the unit.
Measurement values appear too
high or too low.
CAUSES AND SOLUTIONS
Replace both batteries with new ones at the same
time. Check the battery installation for proper
placement of the battery polarities.
Blood pressure varies constantly. Many factors
including stress, time of day, how you wrap the
cuff, may affect your blood pressure.
Review the sections Before Taking a
Measurement and Taking a Measurement.
CARE AND MAINTENANCE
To keep your digital blood pressure monitor in the best condition and protect the unit
from damage, follow the directions listed below:
Do not forcefully bend the cuff. Do not fold tightly.
Clean the monitor with a soft dry cloth.
Do not use any abrasive or volatile cleaners.
Do not attempt to clean the cuff.
CAUTION
Do not submerge the device or any of the components in water. Do not subject
the monitor to extreme hot or cold temperatures, humidity or direct sunlight.
CAUTION
Store the device and the components in a clean, safe location.
CAUTION
Do not subject the monitor to strong shocks, such as dropping the unit on the
floor.
Remove the batteries if the unit will not be used for three months or longer. Always
replace all the batteries with new ones at the same time.
Use the unit consistent with the instruction provided in this manual.
Use only authorized parts and accessories.
Use the unit in a manner consistent with the instructions provided in this manual.
CAUTION
Use only Omron authorized parts and accessories. Parts and accessories
not approved for use with the device may damage the unit.
CAUTION
Changes or modification not approved by the manufacturer will void the user
warranty. Do not disassemble or attempt to repair the unit or components.
TO DELETE ALL VALUES STORED IN THE MEMORY
The values stored in the memory are deleted by User ID. You cannot partially delete
values stored in the memory. All values for the user you select will be deleted.
1. Select your User ID (A or B).
2.PresstheMEMbuttontoturnthemonitoron.
3.WhileholdingtheMEMbutton,presstheSTART/STOPbuttondownformorethan
5 seconds until the display appears as it does in the following illustration.
4. Press the START/STOP button to turn the monitor off.
NOTE: The monitor will automatically turn off after two minutes.
LIMITED WARRANTY
Your BP652 IntelliSense
®
Automatic Wrist Blood Pressure Monitor, excluding the
wrist cuff and batteries, is warranted to be free from defects in materials and
workmanship appearing within 5 years from the date of purchase, when used in
accordance with the instructions provided with the monitor. The above warranties
extend only to the original retail purchaser.
We will, at our option, repair or replace without charge any monitor or wrist cuff
covered by the above warranties. Repair or replacement is our only responsibility and
your only remedy under the above warranties.
To obtain warranty service contact Customer Service by calling 1-800-634-4350 for the
address of the repair location and the return shipping and handling fee.
EnclosetheProofofPurchase.Includealetter,withyourname,address,phone
number, and description of the specific problem. Pack the product carefully to prevent
damage in transit. Because of possible loss in transit, we recommend insuring the
product with return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN
CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS
ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE
IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD
OF THE ABOVE EXPRESS WARRANTY.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER
SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS,
EXPENSES OR DAMAGES.
This warranty provides you with specific legal rights, and you may have other rights
that vary by jurisdiction. Because of special local requirements, some of the above
limitations and exclusions may not apply to you.
FOR CUSTOMER SERVICE
Visit our web site at: www.omronhealthcare.com
Call toll free: 1-800-634-4350
SPECIFICATIONS
Model:
Display:
Measurement Range:
Accuracy/Calibration:
Inflation:
Deflation:
Rapid Air Release:
Measurement Method:
Power Source:
Battery Life:
Operating Temperature
/Humidity:
Storage Temperature
/Humidity/Air Pressure:
Main Unit Weight:
Main Unit Dimensions:
Measurable
circumference of wrist:
Memory:
Contents:
BP652(HEM-6052-Z)
LCD Digital Display
Pressure: 0 to 299 mmHg, Pulse: 40 to 180/min.
Pressure: ±3mmHg or 2% of reading
Pulse: ±5% of reading
Automatic by electric pump
Automatic pressure release valve
Automatic exhaust valve
Oscillometric method
1.5V 2 AAA batteries
Approx. 300 uses with 2 new alkaline batteries
50ºF to 104ºF (10ºC to 40ºC) /15 to 90% RH
-4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH
700-1060 hPa
Approximately 4 oz. (114g)
not including batteries
Approx. 2 3/4"(l) × 2 3/4"(w) × 7/8" (h)
(70 mm × 70 mm × 21 mm) (not including the wrist cuff)
Approximately 5 1/4" to 8 1/2" (13.5 to 21.5 cm)
Up to 100 readings per user
Main Unit, Storage Case, 2
AAA
Alkaline Batteries,
InstructionManual,andQuickStartGuide
NOTE: These specifications are subject to change without notice.
Manufactured for :
OMRONHEALTHCARECo.,Ltd.
Distributed by :
OMRONHEALTHCARE,INC.
1925 West Field Court
Lake Forest, IL 60045 U.S.A.
www.omronhealthcare.com
©2012OMRONHEALTHCARE,INC.
Made in China
Note:
POTENTIALFORRADIO/TELEVISIONINTERFERENCE(forU.S.A.only)
This product has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference
in a residential installation. The product generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions,
may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If the product does
cause harmful interference to radio or television reception, which can be determined by
turning the product on and off, the user is encouraged to try to correct the interference
by one or more of the following measures:
Reorientorrelocatethereceivingantenna.
Increasetheseparationbetweentheproductandthereceiver.
Connecttheproductintoanoutletonacircuitdifferentfromthattowhichthe
receiver is connected.
Consultthedealeroranexperiencedradio/TVtechnicianforhelp.
POTENTIALFORRADIO/TELEVISIONINTERFERENCE(forCanadaonly)
This digital apparatus does not exceed the Class B limits for radio noise emissions
from digital apparatus as set out in the interference-causing equipment standard entitled
DigitalApparatus,ICES-003oftheCanadianDepartmentofCommunications.
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables aux
appareils numériques de Clase B prescrites dans la norme sur le materiel brouilleur:
AppareilsNumériques,ICES-003édictéeparleministerdescommunications.
Changes or modifications not expressly approved by the party responsible for
compliance could void the users authority to operate the equipment.
FCC STATEMENT
USING THE MEMORY FUNCTION
Everytimeyoucompleteameasurementthemonitorstoresthebloodpressureand
pulse rate for each user (A and B) in the memory. The monitor automatically stores up
to 100 sets of measurement values (blood pressure and pulse rate) for each user (A and
B). When 100 sets of measurement values are stored, the oldest record is deleted to
save the most recent values.
AVERAGING FUNCTION
The monitor calculates the average reading based on up to the three most recent sets of
measurement values taken within 10 minutes of the most recent reading.
NOTE: If two sets of measurement values are stored in the memory for the 10 minute
period, the average is based on the two sets of measurement values. If only
one set of measurement values is stored, this is displayed as the average.
TO DISPLAY THE MEASUREMENT VALUES
1.PresstheMEMbuttontodisplaytheaveragereadingbasedonthethreemostrecent
sets of measurement values taken within 10 minutes of the most recent reading. The
Average Value Symbol ( ) appears on the display when the average reading is
shown.
2.PresstheMEMbuttontodisplaythesetofmeasurementvaluesfromthemost
recent to the oldest.
NOTES: The date and time are alternately displayed with the measurement
values.
PresstheMEMbuttonrepeatedlytodisplaythenextvalues.
PressandholdtheMEMbuttontodisplaythevaluesfaster.
3. Press the START/STOP button to turn the monitor off.
CAUTION
Use only 1.5V AAA alkaline batteries with this device.
1. Press the hook on the bottom of the battery cover
and lift the cover off in the direction of the arrow.
2. Install 2 AAA size batteries so the + (positive) and -
(negative) polarities match the polarities of the battery
compartment as indicated.
3. Set the date and time.
Refer to Setting the Date and Time.
4. Replace the battery cover.
NOTE: Make sure that the battery cover is securely in
position.
BATTERY INSTALLATION
INTROD UCTION
Thank you for purchasing the OMRON
®
BP652 IntelliSense
®
Automatic Wrist Blood
Pressure Monitor .
Fill in for future reference.
DATE P URCHASED: __________________
SERIAL NUMBER: ____________________
INSTRUCTION MANUAL
Model B P652
5329415-9A
7 serie s
Automatic Wrist Blood
Pressure M on itor
APPLYING THE WRIST CUFF
CAUTION
Read all of the information in the instruction manual and any other
literature in the box before operating the unit.
CAUTION
This device is intended for use in measuring blood pressure and
pulse rate in the adult population. Do not use this device on infants
or persons who cannot express their intentions.
APPLYING THE CUFF ON THE LEFT WRIST
1. Roll up sleeve. Make sure your sleeve is not rolled up too
tightly on your arm. This may constrict the flow of blood
in your arm.
2. Wrap the cuff directly against your skin.
Do not apply over clothing. Place the cuff over your left
wrist with your left thumb facing upward.
3. Position the cuff leaving a clearance of approximately 1/2
to 1 inch between the cuff and the bottom of your palm.
4. Hold the bottom part of the cuff and wrap it around the
wrist so it fits comfortably and securely around your wrist.
NOTE: Make sure the wrist cuff does not cover the
protruding part of the wrist bone on the outside of
the wrist.
Wrist cuff
Display
Date/Time setting
( ) button
Date/Time setting
( ) button
Battery compartment
Memory (MEM)
button
START/STOP button
User ID selection
switch
SET button
Diastolic blood pressure
Heart guide indicator
Systolic blood pressure
Deflation symbol
Blood pressure level
indicator
Buzzer symbol
Date/Time display
Heartbeat symbol
(Flashes during
measurement)
Memory symbol
(Displayed when viewing
values stored in memory)
Average value symbol
(Refer to
Using the
Memory Function
.)
Irregular heartbeat symbol
Movement error symbol
(Displayed if you move
your body during the
measurement)
User ID symbol (A or B)
(If
GUEST
is selected, A
or B will not be displayed.)
Wrist symbol
Battery low symbol
Pulse display
User A User B
ERROR INDICATORS
SYMBOL CAUSE
Device error.
CORRECTION
Keep still and take another
measurement.
Replace the batteries.
Refer to Battery Installation.
Contact Customer Service.
Remove the monitor from your
wrist. Wait 2-3 minutes and then
take another measurement. Repeat
the steps in section Applying the
Wrist Cuff. If this symbol
continues to appear, contact your
doctor.
Batteries are worn.
Irregular or weak
pulses are detected.
Movement during
measurement.
Monitor could not
detect pulse wave.
Cuff over-inflated.
Cuff under-inflated.
Cuff not applied
correctly.
Take another measurement and
remain still until the measurement
is complete.
Remove the monitor from your
wrist.
Read Taking a Measurement.
Wait 2-3 minutes.
Take another measurement.
Alternating date/time
display
NOTE: The monitor will automatically turn off after two minutes.
6. Press the START/STOP button to turn the monitor off.
5. After the monitor has detected your blood pressure and
pulse rate, the cuff automatically deflates. Your blood
pressure and pulse rate are displayed.
NOTE: To stop the inflation or measurement, push the START/STOP button. The
monitor will stop inflating, start deflating, and will turn off.
The cuff starts to inflate automatically.
As the cuff inflates, the monitor automatically determines your ideal inflation level.
This monitor detects your blood pressure and pulse rate during inflation.
The Heartbeat Symbol ( ) flashes at every heartbeat.
Remain still and do not move until the entire measurement process is completed.
IftheHeartGuideIndicatorissettoon2,theorangelight
will turn to blue when your arm is in the correct position.
(If set to either on1 or oFF, the cuff will inflate
automatically.)
4. Hold your arm across your chest at heart level.
Hold your elbow firmly to avoid body movement.
Sit still and do not talk or move during the measurement.
All symbols appear on the display.
3. Press the START/STOP button.
2. Select your User ID (A or B).
The monitor is designed to take measurements and store the measurement values in
memoryfortwopeopleusingUSERIDAandUSERIDB.Themonitorcanbeusedto
takeasinglemeasurementusingtheGUESTmodearenotstoredinthememory.
1. Sit comfortably on a chair with your feet flat on the floor.
User A User B
Loading ...
Loading ...
Loading ...