6. Saving, retrieving and deleting results
The results of each successful measurement are stored
together with date and time. With more than 30 items of
measured data, the earliest items of data measured are lost.
Use the SET button and then the -/+ buttons to select the
desired user memory and confirm using the button
. Pres-
sing the M+ button displays the average value AVG of the 3
most recent measurements stored in the user memory. By
pressing the + button, the most recent measurements are
displayed; by pressing the - button, the oldest are displayed.
The average value AVG can only be displayed if there are 3
saved values in the respective user memory.
To clear the memory, you must first select a user memory.
Hold the SET button pressed for approx. 3 seconds. All the
values in the current user memory are deleted after 3 brief
beeps are output.
If you wish to change the user memory, see “Select memory”.
7. Cleaning and storing the instrument
Clean your device and cuff carefully only with a slightly mois-
tened cloth.
Do not use detergents or solvents.
On no account must you immerse the computer in water,
otherwise liquid can enter it and cause demage.
When storing the device, make sure that no heavy objects
are placed on top of it. Remove the batteries. The cuff tube
should not have any sharp kinks.
8. Error messages / troubleshooting
In case of faults, the Err message appears in the display.
Error messages can occur when
1. inflation pressure is higher than 300 mmHg,
2. blood pressure is unusually high or low,
3. you move or talk during the measurement.
In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do
not move or talk. Re-insert the batteries if necessary, or else
replace them.
9. Specifications
Model no. SBM 21
Measurement
method
Oscillometric, non-invasive blood pres-
sure measurement on the upper arm
Measurement range Cuff pressure 0 300 mmHg,
systolic 60 – 280 mmHg,
diastolic 30 – 200 mmHg,
Pulse 30 – 180 beats/minute
Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
inaccuracy
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory 4 x 30 memory spaces
Dimensions L 155 mm x W 110 mm x H 70 mm
Weight Approx. 395 g (without batteries, with
cuff)