Recommended separation distances between portable and mobile RF
communications equipment and the EBP-UA5 Upper-Arm Blood Pressure Monitor
The EBP-UA5 Upper-Arm Blood Pressure Monitor is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the EBP-UA5 Upper-Arm Blood Pressure Monitor
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the EBP-UA5 Upper-Arm Blood Pressure Monitor as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz 80 MHz to 800 MHz 80 MHz to 2.7 GHz
0.01 N/A 0.035 0.07
0.1 N/A 0.11 0.22
1 N/A 0.35 0.7
10 N/A 1.11 2.21
100 N/A 3.5 7