25
Interference may occur in the vicinity of
equipment marked with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
A. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the EBP-UA5 Upper-Arm Blood Pressure Monitor is
used exceeds the applicable RF compliance level above, the EBP-UA5 Upper-Arm Blood Pressure Monitor should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the EBP-UA5 Upper-Arm Blood Pressure Monitor.
B. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.