Guidance and Manufacturer’s Declaration – Electromagnetic Immunity –
for all EQUIPMENT and SYSTEMS
The EBP-UA5 Upper-Arm Blood Pressure Monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the EBP-UA5 Upper-Arm Blood Pressure Monitor should assure that it is used in such an
environment.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply lines N/A
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV differential mode N/A
Mains power quality should be that
of a typical commercial or hospital
environment.