Loading ...
Loading ...
Loading ...
50
1. Electromagnetic Compatibility Information
Guidance and manufacture’s declaration – electromagnetic emissions – for all EQUIPMENT and SYSTEMS
Guidance and manufactures declaration – electromagnetic emission
The
SBM 38
blood pressure monitor is intended for use in the electromagnetic environment specified below. The customer of the user of the
SBM38
blood pressure monitor should assure that it is used in such and environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1 The
SBM 38
blood pressure monitor uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic equipment.
RF emission CISPR 11 Class B
The
SBM 38
blood pressure monitor is suitable for use in all establishments, including domestic establish-
ments and those directly connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Guidance and manufacture’s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The
SBM 38
blood pressure monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the
SBM 38
blood pressure monitor should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floor are covered with synthetic
material, the relative humidity should be at least 30%.
Power frequency (50Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical loca-
tion in a typical commercial or hospital environment.
Loading ...
Loading ...