Electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
The Blood pressure monitor (SBM 22) is intended for use in the electromagnetic environment specified below. The customer
of the user of the Blood pressure monitor (SBM 22) should assure that it is used in such and environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If floor are
covered with synthetic material, the relative humidity should be
at least 30%.
Power frequency
(50Hz) magnetic
field
IEC 61000-4-8
3A/m 3A/m Power frequency magnetic fields should be at levels characte-
ristic of a typical location in a typical commercial or hospital
environment.
Guidance and manufacturer’s declaration – electromagnetic immunity for EQUIPMENT and SYSTEMS that are not
LIFE-SUPPORTING
The Blood pressure monitor (SBM 22) is intended for use in the electromagnetic environment specified below. The customer
or the user of Blood pressure monitor (SBM 22) should assure that it is used in such an environment.