7. Saving, displaying and deleting measured
values
The results of every successful measurement are stored
together with the date and time. If there are more than 30
measurements, the oldest measurements are lost.
Press the memory button M. Select the desired user
memory (
... ) by pressing the memory button M
again.
If you press the function button +, the average value of
all the stored measured values in the user memory will be
displayed. If you press the function button + again, the
average value of the morning measurements for the last 7
days will be displayed (morning: 5 a.m. – 9 a.m., display
). If you press the function button + again, the average
value of the evening measurements for the last 7 days will
be displayed (evening: 6 p.m. – 8 p.m., display
). If you
continue to press the function button +, the most recent
individual measured values are displayed in turn with the
date and time.
To switch off, press the START/STOP button
.
If you forget to switch off the device, it will switch itself off
automatically after 30 seconds.
If you want to delete the whole memory for a specific
user, press the memory button M. Press and hold down
the memory button M and the setting button SET simulta-
neously for 5 seconds.
8. Error messages/troubleshooting
In the event of errors, the error message _ appears on the
screen.
Error messages may appear if:
It was not possible to correctly record the pulse:
1
;
No measurement could be taken:
2
;
The cuff is fastened too tightly or loosely:
3 ;
Errors occur during the measurement:
4
;
The pump pressure is higher than 300 mmHg:
5
;
There is a system error. If this error message appears,
please contact Customer Services:
6
.
The batteries are almost empty: .
In such cases, repeat the measurement. Ensure that you do
not move or speak.
If necessary, reinsert or replace the batteries.
9. Cleaning and storing the unit
Clean your blood pressure monitor carefully using a
slightly damp cloth only.
Do not use any cleaning agents or solvents.
Under no circumstances should you hold the device
under water, as this can cause liquid to enter and damage
the device.
If you store the unit, no heavy objects should be placed
on top of it. Remove the batteries. The cuff line should not
be bent sharply.