MKF / MKFT (E5) 06/2020 page 80/176
Condition
Information
icon
Text information
Start after
condition occurred
Operation line “Compr. air dryer” on. Option-
al compressed air dryer activated.
Turn operation line off when not using the
compressed air dryer!
“Compressed Air
Dryer”
immediately
Notifications are not shown in the event list.
11.1.2 Alarm messages
Condition Alarm message
Start after
condition occurred
Chamber door open “Door open after 5 minutes
Exceeded setpoint of the safety controller class 2 “Safety controller immediately
Exceeded maximum or minimum temperature (option
temperature safety device class 2)
“Temperature safety
device”
immediately
Temperature sensor defective
e.g. “- - - - or
“<-<-<” or “>->->”
immediately
Safety controller temperature sensor defective
“Safety controller
sensor”
immediately
MKF/MKFT 115, 240, 720:
Fault in refrigerating machine. Contact BINDER service.
Compressor
overcurrent
immediately
Alarm messages are displayed in the list of active alarms until acknowledging them. They are also shown
in the event list.
11.1.3 Messages concerning the humidity system
Condition and measures Message
Start after condition
occurred
The humidity module is defective. Contact BINDER
service
“Humidity system immediately
MKF 56:
The humidity module cannot fill up.
In case of freshwater supply via water pipe:
The water tap is closed, or the chamber is defective
(e.g. inlet valve of humidity module).
In case of freshwater supply via external freshwater can
(option):
Water can is empty. Humidification is turned off. In case
of refrigerating operation, the interior is strongly dehu-
midified. When the water supply is functional again, the
humidity system restarts, or the chamber is defective.
“Freshwater supply” immediately
MKF/MKFT 115, 240, 720:
Freshwater can is too empty to allow normal function, or
the floating switch is defective.
The humidification system turns off.
Fill up the water can or open the water supply.
During the one-hour preheating phase: message without
significance.
Freshwater can
empty
after 60 sec.