MKF / MKFT (E5) 06/2020 page 54/176
When entering a value outside the setting range, the message: “Value outside of limits! (Min:
xxx, Max: xxx)” appears (xxx is a wildcard for the limits of the respective parameter). Press the
Confirm icon and repeat the entry with a correct value.
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press
the Close icon to exit the menu without taking over the entries.
7.2 Direct setpoint entry via Normal display
Alternatively you can also enter the setpoints directly via Normal display.
Normal display (example: MKF 56).
Select the setpoint you want to change.
Example: “Temperature” entry menu.
Enter the desired setpoint and confirm entry
with Confirm icon
7.3 Special controller functions via operation lines
Press the Setpoint setting icon to access the “Setpoint” setting menu from Normal display.
You can define the switching state of up to 16 operation lines (control contacts). They are used to activate
/ deactivate special controller functions.
Operation line Humidity offserves to turn off the humidity.
Operation line Idle modeactivates / deactivates the operating mode Idle mode”.
Operation lines Switching output 1” up to “4” can be used to turn on and off any equipment connected
to the zero-voltage switching outputs (DIN sockets (7) and (8) (chap. 20).
Operation line Compr. air dryerserves to activate the compressed air dryer (option, chap. 21.6).
Operation line “Compressed air valve” serves to open the solenoid valve of the compressed air con-
nection (for options compressed air connection, chap. 21.5, or compressed air dryer, chap. 21.6)
Operation line Anti-condensationserves to switch the anti-condensation protection (chap. 19).
The other operation lines are without function.