MKF / MKFT (E5) 06/2020 page 49/176
5.7 Performance during and after power failures
During a power failure, all controller functions are shut down.
After the power returns, all functions return to the same status the chamber had before power failure. The
controller continues to function in the original operating mode it was in previously before the power failure
occurred.
Performance after power failure in Idle mode
Control is deactivated
Performance after power failure in Fixed value operation mode
All functions return to the same status the chamber had before power failure. The set-points are im-
mediately resumed.
Performance after power failure during time program operation
The program is resumed at the point where the interruption occurred with the latest set-points reached
during the program run.
Performance after power failure during week program operation
The week program continues with the values corresponding to the current time.
Power failure and power return are noted in the event list (chap. 15.3).
If during power failure an alarm has occurred (tolerance range, safety controller, optional over/under tem-
perature safety device class 2), confirm the alarm. See chap. 11.3.
5.8 Performance when opening the door
MKF 56: Immediately after opening the door, the fan runs at the minimum speed (30%). 60 seconds after
opening the door, the fan turns off.
MKF/MKFT 115, 240, 720: The fan turns off immediately after opening the door.
After 60 seconds from opening the door, heating, refrigeration, humidification, dehumidification and fan
turn off. The compressor continues operating during 5 minutes without cooling function.
After closing the door, heating, refrigeration, humidification, dehumidification and fan turn on again.