MKF / MKFT (E5) 06/2020 page 116/176
17. Humidification / dehumidification system
The chamber is equipped with a capacitive humidity sensor. This results in a regulatory accuracy of up to
+/- 2.5 % r.h. of the set point. The temperature-humidity diagrams (Figure 19) show the possible working
range for humidity.
In the setpointsmenu you can turn humidity control (humidification and dehumidification) on or off
with the setting Control on/off(chap. 6.3).
With humidity control turned off, the humidification module cools down. After activation it will take up to
20 minutes until the humidification function is fully available again. This setting is required when
operating the chamber without a water connection in order to avoid humidity alarms.
Operation line Humidity offserves to turn off the humidification / dehumidification system in Fixed
value operation (chap. 7.3, time program operation (chap. 9.7.3) and week program operation (chap.
10.6.5). This allows configuring the disconnection for individual program sections.
When the humidification / dehumidification system is turned off via operation line it remain on standby
(filled and heated). Therefore it is immediately available after turning on.
When operating the chamber with activated humidity, humidity control turns off automatically at tempera-
ture set-points below 0 °C / 32 °F or above 95 °C / 203 °F. The information icon Humidity offis dis-
played in the screen header in Normal display. When the temperature setpoint is set back to the range
from 0 °C / 32 °F to 95 °C / 203 °F, humidity control turns on again and the information icon “Humidity off”
disappears.
0
10
20
30
40
50
60
70
80
90
100
10
20
30
40
50
60
70
80
90
100
% r.H.
°C
A
B
0
10
20
30
40
50
60
70
80
90
100
10
20
30
40
50
60
70
80
90
100
% r.
H.
°C
A
B
C
Regular chambers Chambers with optional compressed air dryer
Figure 19: Temperature-humidity diagrams
Range A: Control range of temperature and relative humidity.
Range B: Discontinuous range (no continuous operation, max. 24 hours. Observe hints on defrosting,
chap. 18)
Range C: Enlarged climatic range with optional compressed air dryer.
Loading ...
Loading ...
Loading ...