21
Printing the Measurement Data
When the [PRINT] switch on the back of the monitor is set to
"ON", the measurement results will be printed out when the
measurement is complete.
The items that appear in the printed results depend on the
settings of function selection mode.
("Function selection mode" page 23)
Example of printed measurement results
(1) Measurement results
The results of the measurement.
"Mean arterial pressure (MAP)" will appear
only when "MAP print" is set to "ON" in
function selection mode.
(2) Arm position
If measurement was started with the elbow
detection lamp remaining lit on the display,
"Measurement was performed at the correct
arm position" will be indicated on the printed
measurement results.
If the elbow detection lamp was still blinking
when measurement was started, "Placing
your elbow on the elbow rest will enable a
more accurate measurement" will be indicated
on the printed measurement results.
(3) Noise level/oscillometric envelope
If the user moves or talks during the
measurement, oscillations similar to pulses
will occur at the measuring site. The monitor
interprets these oscillations as noise.
If there is too much noise, the measurement
will not be accurate. In this case, repeat
measurement, taking care not to move or talk.
(4) Results commentary
A comment will appear based on the
measurement results.
(5) Advertisement print
If "Image upload" in function selection mode is
used to upload an advertisement image from
an SD memory card*, the image will be
printed.
*For a compatible SD memory card, please contact your local
authorized OMRON dealer/supplier.
Print pattern: A
Excellent measurement results
Determined systolic blood pressure (SYS)
Determined diastolic blood pressure (DIA)
For these measurement results,
the oscillometric envelope is a
clean mountain shape, there is
no noise during the measurement
such as motion artifacts, and the
results are highly reliable.
Measurement results with low reliability
Signal that is assumed to be noise
Assumed systolic blood pressure
For these measurement results,
the oscillometric envelope is not
a clean mountain shape and it is
likely that there was noise and it
is conceivable that there was
noise such as motion artifacts
during the measurement.
(1)
(2)