Binder MK 115 -liter dynamic climate chambers for rapid temperature changes

User Manual - Page 120

For MK 115.

PDF File Manual, 163 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
MK / MKT (E5) 08/2020 page 120/162
Fault description
Possible cause
Required measures
Controller (continued)
Controller does not equilibrate to
setpoints entered in Fixed value
operation mode
Controller is not in Fixed value
operation mode.
Change to Fixed value operation
mode.
Controller does not equilibrate to
program set-points.
Controller is not in program
operation mode, or program
delay time is running.
Start the program again. If appropri-
ate, wait for the program delay time.
Program duration longer than
programmed.
Tolerances have been pro-
grammed.
For rapid transition phases, do NOT
program tolerance limits in order to
permit maximum heating, speed.
Program keeps the last program
setpoint constant while in setting
“ramp”.
Program line with setting
“ramp” is incomplete.
When programming with setting
“ramp”, define the end value of the
desired cycle by adding an addi-
tional section with a section time of
at least one second.
Ramp temperature transitions
are only realized as steps.
Setting “step” has been se-
lected.
Select setting “ramp”.
Acknowledging the alarm does
not cancel the alarm state.
Cause of alarm persists.
Remove cause of alarm. If the alarm
state continues, contact BINDER
service.
Alarm message: “- - - - “ or
“<-<-<“ or “>->->“
Sensor rupture between sen-
sor and controller or Pt 100
sensor defective.
Contact BINDER service.
Short-circuit.