MK / MKT (E5) 08/2020 page 111/162
Note:
Automatic regeneration in a 2-minute interval is required for the function of the compressed air dryer. As
this happens, a small amount of compressed air is suddenly released into the environment, what is asso-
ciated with some noise.
The desiccant cartridges must be replaced after no more than 17,500 operating hours or 2 years. The filter
must be replaced after a maximum of 8,760 hours or 1 year. We recommend an annual maintenance inter-
val.
An exchange is also necessary when the status display on the rear panel is lit or flashes in yellow. Check
the status display approx. once per month.
Normal operating state
Filter elements and desiccant cartridges must be replaced. Contact BINDER Service
Alarm. Contact BINDER Service
Status display
Figure 18: Rear chamber view with optional compressed air dryer (example: MKT 115)