Binder KBF-S 1020 Microclimate cabinet

User Manual - Page 72

For KBF-S 1020. Series: KBF-S

PDF File Manual, 91 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
KBF-S / KBF-S-UL (E6) 02/2019 page 72/91
24.3 Definition of usable volume
The usable volume illustrated below is calculated as follows:
c
c
b
a
C
B
A
b
a
A, B, C = internal dimensions (W, H, D)
a, b, c = distance to wall
a = 0.1*A
b = 0.1*B
c = 0.1*C
V
USE
= (A - 2 * a) * (B - 2 * b) * (C - 2 * c)
Figure 20: Determination of the useable volume
The technical data refers to the defined usable volume.
Do NOT place samples outside this usable volume.
Do NOT load this volume by more than half to enable sufficient airflow inside the chamber.
Do NOT divide the usable volume into separate parts with large area samples.
Do NOT place samples too close to each other in order to permit circulation between them and
thus obtain a homogenous distribution of temperature and humidity.
24.4 KBF-S / KBF-S-UL technical data
Chamber size
115
240
720
1020
Exterior dimensions
Width, net
mm / inch
880 / 34.65
925 / 36.42
1250 / 49.21
1250 / 49.21
Height, gross (incl. feet/castors)
mm / inch
1050 / 41.34
1460 / 57.48
1925 / 75.79
1925 / 75.79
Depth, net
mm / inch
650 / 25.59
800 / 31.50
890 / 35.04
1145 / 45.08
Depth, gross (including door handle, I-
triangle, connection and 30 mm for
cable)
mm / inch 730 / 28.74 880 / 34.65 970 / 38.19 1230 / 48.43
Wall clearance rear (minimum)
(spacer)
mm / inch 100 / 3.94 100 / 3.94 100 / 3.94 100 / 3.94
Wall clearance side (minimum)
mm / inch
160 / 6.29
160 / 6.29
160 / 6.29
160 / 6.29
Doors
Quantity of doors
1
1
2
2
Interior dimensions
Width
mm / inch
600 / 23.62
650 / 25.60
973 / 38.31
973 / 38.31
Height
mm / inch
483 / 19.02
785 / 30.91
1250 / 49.21
1250 / 49.21
Depth
mm / inch
351 / 13.82
485 / 19.09
576 / 22.68
836 / 32.91
Interior volume
l / cu.ft.
102 / 3.6
247 / 8.7
700/ 24.7
1020 / 36.0
Steam space volume
l / cu.ft.
156 / 5.5
348 / 12.3
918 / 32.4
1280 / 45.2
Racks
Quantity of racks (regular)
2
2
2
2
Quantity of racks (max.)
5
9
15
15
Maximum load per rack
kg / lbs.
30 / 66
30 / 66
45 / 99
45 / 99
Maximum permitted total load
kg / lbs.
100 / 220
100 / 220
150 / 331
150 / 331
Loading ...