KBF-S / KBF-S-UL (E6) 02/2019 page 56/91
16.5 Error during data transmission
In the event of an error, the message ERR (error) is displayed.
Read error (example).
Import configuration.
16.6 Removing the USB stick
Logging off the USB stick is not possible / required.
Be sure that no data recording is running (chap. 16.4).
After removing the USB stick, the controller is back in the same menu as before when connecting the
USB stick.
17. Self-test function
The self-test function enables an automated check of the proper chamber functioning as well as a target-
ed and reliable fault analysis. In this case, the chamber successively undergoes various defined operating
states, which serves to determine reproducible characteristic values. These characteristic values provide
information on the performance and precision of the individual functional systems of the chamber (e.g.,
heating, refrigeration, humidification) of the chamber.
The results of the self-test are stored in the service recorder of the controller. You can export them using
the controller’s USB interface and send them to BINDER Service (use function “Export recorder data” to
USB stick, chap. 16). BINDER Service will evaluate the data using an analyzing tool.
In order to allow an optimum comparison of the determined characteristic values with the
reference characteristic values, the ambient temperature should be in the range of +22 °C +/-
3 °C / 71.6 °F +/- 5.4 °F.
The chamber shall be unloaded (empty with standard equipment).
17.1 Activating the self-test function
Required access level: „Admin“.
Path: Normal display
Settings Various
Self-test
Press the OK button to enable the settings.
Following, you can choose between the following settings:
“0” (function off)
“1” (reduced duration: 12-18 hours)