12. Tolerance range settings
In this menu you can define for temperature and humidity the deviation between the actual value and set-
point of, which that shall cause a tolerance range alarm. The entered value defines the limit of permitted
deviations from the set-point (Exceeding and falling below). Reaching this limit triggers tolerance alarm.
In addition you can specify a delay time for this alarm.
If there are any actual values outside the tolerance range, after the configured alarm delay time the alarm
message “Temperature range” and / or “Humidity range” is displayed in Normal display (chap. 13.1). If the
alarm buzzer is activated (chap. 13.2) there is an audible alert.
This function only activates after the set-point has been reached once.
Required access level: Admin”.
12.1 Setting the delay time for tolerance range alarm
Path: Normal display Settings Various Range alarm delay
(min)
Press the OK button to enable the setting.
Setting the tolerance range alarm delay.
The current setting flashes Use the arrow buttons to enter the desired
time after which the range alarm shall be triggered. Entry range: 1 up to
300 minutes.
Confirm the entry with the OK button.
Range alarm delay (m
With the arrow-down button you can now change to the temperature tolerance range setting.
With the Back button you can go back to the “Varioussubmenu and, repeatedly pressing it, to Normal
display.
12.2 Setting the temperature tolerance range
Path: Normal display Settings Various Temperature range
Press the OK button to enable the setting.
Setting the temperature tolerance range
The current setting flashes. Enter the desired temperature range with
the arrow buttons. Entry range: 1,0 °C up to 10,0 °C
Confirm the entry with the OK button.
Temperature range
With the arrow-up button you can go back to the tolerance range alarm delay setting.
With the arrow-down button you can now change to the humidity tolerance range setting.
With the Back button you can go back to the “Varioussubmenu and, repeatedly pressing it, to Normal
display.