The Monitor uses the osclllometric method of blood pressure measurement. Measurement Automatic Electronic Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate, with an air wrist cuff buckled around one’s wrist according to the instructions in the “ ATTACHING THE WRIST CUFF.” The expected life of the product is 5 years.
The product complies with the electromagnetic compatibility requirement of EN60601-1-2 and safety standards of EN60601-1 and performance of IEC 80601-2-30 as specified in EEC directive 93/42/EEC.
Make sure there is no connection tubing kinking before start measuring to avoid any injury to patient.
For any patient, do not measure more than 3 times continuously, it should be at least above 5 minutes of interval rest between any two measurements, otherwise will cause extravasated blood.
Do not measure your blood pressure over 6 times each day.
Do not apply the cuff over a wound as this can cause further injury.
Do measure on the wrist which is on the side of a mastectomy, otherwise it could cause injury.
Observe the air pressure value from the LCD display.
When measuring, it could not exceed 299 mmHg, otherwise Please press “on/off ” button to stop_
Do not use force to bend the wrist cuff or the air tube.
Do not knock or drop the main unit.
Always use the specified accessories in the manual, the use of other parts not approved by the manufacturer may cause faults or injuries.
For service information, parts list etc., please contact the dealer.
Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands.
Blood pressure is measured in millimeters of mercury (mmHg). One's natural blood pressure is represented by the fundamental pressure, which is measured first thing in the morning while one is still at rest and before eating.
Hypertension, an abnormally high arterial blood pressure, if left unattended, can cause many health problems including stroke and heart attack.
Hypertension can be controlled by altering lifestyle, avoiding stress and with medication under a doctor's supervision.
To prevent hypertension or keep it under control:
Blood pressure measured at a clinic or doctor's office may cause apprehension and produce an elevated reading, 25 to 30 mmHg higher than that measured at home, Home measurement reduces the effects of outside influences on blood pressure readings, supplements the doctor's readings and provides a more accurate, complete blood pressure history.
Standards for assessment of high blood pressure, without regard to age, have been established by the World Health Organization (WHO), and shown in chart below.

An individual's blood pressure varies greatly on a daily and seasonal basis. It may vary by 30 to 50 mmHg due to various conditions during the day. In hypertensive individuals, variations are even more pronounced.
Normally, the blood pressure rises while at work or play and falls to its lowest levels during sleep. So, do not be overly concerned by the results of one measurement.
Take measurements at the same time every day using the procedure described in this manual, and know your normal blood pressure. Many readings give a more comprehensive blood pressure history.
Be sure to note date and time when recording your blood pressure doctor to interpret your blood pressure data.

1. If you are taking medication, consult with your doctor to determine the most appropriate time to measure your blood pressure. NEVER change a prescribed medication without first consulting with your doctor.
2. For people with irregular or unstable peripheral circulation problems due to diabetes, liver disease, hardening of the arteries, etc., there may be fluctuation in blood pressure values measured at the upper arm versus at the wrist.
3. Measurements may be impaired if this device is used near televisions, microwave ovens, X-ray, mobile phone equipment or other devices with strong electrical fields. To prevent such interference, use the monitor at a sufficient distance from such devices or turn them off.
4. Before using, should wash your hands.
5. Do not measure on the arm which simultaneously used monitoring ME Equipment, otherwise it could cause loss of function.
6.Consult your doctor if the unexpected readings are obtained, also please refer to “Trouble shooting” of the manual.
7. The reading is probably a little lower than measured in the hospital due to the steady mood at home.
8. Cuff pressure range 0-299mmHg
1. Memory can store 90 measurements.
2. Large and clear LCD display.
3. WHO blood pressure classification display.
4. Easy to use, Press a button to automatically measure, record the measurement values and measurement time.
5. Automatically turns off (within 1 minute) to save power.

1. Remove the battery cover.
2. Insert new batteries into the battery compartment as shown, taking care that the polarities(+) and (-) are correct.
3. Close the battery cover, Use only LR03, AAA batteries.


Disposal of empty battery to the authorized collecting party subject to the regulation of each individual territory.
CAUTION:
(LOW BATTERY mark) blinks in the display, replace all batteries with new ones. Do not mix old and new batteries. It may shorten the battery life, or cause the device to malfunction.
(LOW BATTERY mark) does not appear when the batteries run out.
If you are not going to use the unit for a long period of time (approximately three months or more), remove the batteries.
Replace worn batteries with their polarities in the correct direction.

1. Press “SET” key to turn on.
2. Press and hold “SET” key until the year number displays and flashes on LCD to enter setting mode.
3. Press “MEM” key to adjust the year, then press “SET” key again to save your setting and enter the month setting mode.
4. Press “MEM” key to adjust the month. Following the same steps to adjust date/hour/ minute/(Voice on/off) until setting completed(“
” is the On, “
” is the Off) Non-talking model does not have this function, or you may choose method No.2 to set the voice

Method No.2 for voice setting: Please press and hold the“MEM” button when the device is in shutdown state to enter the voice setting, please press the “MEM” button again to choose “ON” or “OF”, press “ON/OFF” button to quit.


The goods have mm Hg(mmHg), kPa (kPa) two kinds of blood pressure display units(mmHg factory to express).
Boots continued to press the ON/OFF button exceeding five seconds.
The units will be chosen by the above shows mmHg/kPa after decontrol, After the nomal boot unit values are shown as blood pressure.
Also select memory unit value changes.
Diastolic blood pressure Reference material: journal of hypertension 1999. vol 17 No.2

1. Fastening the wrist cuff
1) Wrap the wrist cuff around your wrist about (1-2)cm above your hand as shown in the figure below.

2) Fasten the wrist cuff tightly by using the Velcro Strip.
For proper measurements, fasten the wrist cuff tightly and measure on a bare wrist.
2. How to take proper measurements

1. Fasten the wrist cuff according to the instructions in “ATTACHING THE WRIST CUFF.”
2. Press the “ON/OFF” button. All icons appear two seconds on DISPLAY, then switch to measurement, and display “0”or last measurement record.

3. Start measurement, the cuff in the strap will automatically inflate.
The mark will flash on LCD. When complete, the results will be displayed.

READ MEMORY
Press “MEM” button, a memory reading out the latest measurements ,“MEM” for the buttons(UP). “SET” button for the memory (DOWN)
Power Measurement closure or after the end of the state .can press the “MEM” button read out the latest measurement of memory.
DELETE MEMORY
The state read out the memory press the (memory) button five seconds, the LCD display “
” has been to delete all memory.

To keep your digital blood pressure monitor in the best condition and protect the unit from damage, follow the directions listed below:
Keep the monitor in the storage case when not in use.
Do not fold the arm cuff too tightly.
The fabric fastener could touch the inner surface of the arm cuff and damage it.
Clean the monitor and cuff with a soft dry cloth. Do not use any abrasive or volatile cleaners.

CAUTION
(Pressure calibration is done by EU representatives)
Measuring Method | ||
ERROR DISPLAY | POSSIBLE CAUSE | HOW TO CORRECT |
Nothing is displayed When you push the POWER button or | No battery installation | Insert batteries |
Battery worn out | Replace new batteries | |
The polarities of batteries placed wrongly | Insert battery in the correct polarities | |
E1:can’t normally Increase pressure | Check your wrist cuff if any air leakage | Replace wrist cuff with new one |
E3 inflate pressure too high |
| Re-measurement or send back dealer for re-calibrate pressure |
E2E4:have shaking while measurement | Hand or body shaking while measurement | keeping static and correct gesture to measure again |
| Battery low power | Replace battery and measure again |
The systolic pressure Value or diastolic Pressure value too high | 1.The wrist cuff was held lower than your heart | keeping correct position and gesture to measure again |
2.The wrist cuff was not attached properly | ||
3.You moved your body or spoke during measurement | ||
The systolic pressure Value or diastolic Pressure value too low | 1 .The wrist cuff was held higher than your heart | |
2.you moved your body or Spoke during measurement |