8
all test pieces (Test Base and Sample Receiver) must
be removed and discarded. The instrument will
proceed to the Home screen. Press Run Test and
restart the test using a new Test Base and Sample
Receiver.
Wait for the Sample Receiver to Warm Up. Do not
remove the Sample Receiver from the instrument
once Warm Up begins.
Caution: DO NOT REMOVE THE FOIL SEAL
UNTIL PROMPTED BY THE INSTRUMENT. DO
NOT close the lid or insert the sample until prompted by
the instrument.
Step 4
Direct Nasal, Throat or Nasopharyngeal Swab Test
Procedure
When prompted, remove the foil seal and place the
patient swab to be tested into the Sample Receiver.
Mix the swab in the liquid for 10 seconds. This helps
remove the sample from the swab. Lift the swab out of the
liquid and press the swab head against the side of the
Sample Receiver to remove excess liquid. Once the swab is
removed, touch ‘OK’ to proceed.
Discard the swab into a biohazard waste container.
Caution: To ensure that the Sample Receiver
remains in the instrument while removing the foil
seal, place two fingers along the outer edge of the
Sample Receiver to hold it in place. If the Sample
Receiver spills after warm up, cancel the test by
pressing the Home button. Remove and discard the
test pieces (Sample Receiver and Test Base) and
clean the instrument. Press Run Test to start a new
test using a new Test Base and Sample Receiver.
Step 5a
Press the White Transfer Cartridge into the Blue
Sample Receiver
Listen for a click.
When the Transfer Cartridge is properly attached to the
Sample Receiver, the orange indicator on the Transfer