255
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electro-
magnetic site survey should be considered. If the measured field strength in the
location in which the UNI-DEX is used exceeds the applicable RF compliance level
above, the UNI-DEX should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-ori-
enting or re-locating the UNI-DEX.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
Recommendedseparationdistances
Recommended separation distances between portable and mobile RF
communicationequipmentandtheUNI-DEX
The UNI-DEX is intended for use in the electromagnetic environment in
whichRFdisturbancesarecontrolledThecustomerortheuseroftheUNI-
DEXcanhelppreventelectromagneticinterferencebymaintainingamini-
mumdistancebetweenportableandmobileRFcommunicationsequipment
(transmitters)andtheUNI-DEXasrecommendedbelowaccordingtothe
maximumoutputpowerofthecommunicationsequipment
Rated maximum
output power of
transmitter
W
Separation distance according
to frequency of transmitter
m
150 kHz to
80 MHz
d = 1.2
√P
80 MHz to
800 MHz
d = 1.2 √P
800 MHz to 2.5 GHz
d = 2.3
√P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23