The result of the control solution test should be within the range
printed on the test strip vial la bel. If the test result falls outside the
specied range, repeat the test, carefully following the steps
above.
[2] When the meter is connected to your iHealth Gluco-Smart App
on your mobile device:
Step l: Launch the app.
Step 2: Tap the CTL button to switch to control (CTL) mode.
Step 3: Follow Steps 3-4 above in section [1].
Step 4: The control solution test result will appear on the app on
your mobile device.
Out-of-Range Results
Results falling outside the specied range may be caused by:
. An error in the test
. Expired or contaminated control solution
. An expired or contaminated test strip
. Meter malfunction
If you continue to get control solution test results that fall outside
of the range printed on the vial, the meter may not be working
properly. Discontinue use and call iHealth Labs Customer Service
at 1-855-816-7705 for assistance or email at
Control solution is available for purchase directly on the iHealth
website store at www.ihealthlabs.com and other select retailers.
NOTE:
. Do not use expired control solution.
. The control solution range printed on the test strip vial is for
control solution use only. It is not a recommended range for your
blood glucose level.
19