Note: Be sure to set the meter and/or app on the CTL mode before
performing a control solution test. The control solution test result
will not be saved in the meter.
Step 2: Press the memory button again to turn o the CTL mode
and switch back to the regular testing mode.
Step 3: Apply the control solution.
. Shake the control solution vial before each use.
. Squeeze a drop of control solution into the vial cap. For better
results, it is recommended that you use the second drop of the
control solution (discard the rst drop).
. Hold the meter and move the absorbent hole of the test strip to
absorb the drop. Once the conrmation window lls completely,
the meter will start counting down. Remove the control solution
sample from the test strip when the countdown begins.
Note: To avoid contaminating the entire vial of control solution, do
not directly apply control solution onto a strip.
Step 4: Read and compare the results.
After the meter counts down to "1"; the control solution test result
will appear on the meter display.
18