Note: Be sure to set the meter and/or app on the CTL mode before
performing a control solution test. The control solution test result
will not be saved in the meter.
Step 2: Press the memory button again to turn o the CTL mode
and switch back to the regular testing mode.
Step 3: Apply the control solution.
. Shake the control solution vial before each use.
. Squeeze a drop of control solution into the vial cap. For better
results, it is recommended that you use the second drop of the
control solution (discard the rst drop).
. Hold the meter and move the absorbent hole of the test strip to
absorb the drop. Once the conrmation window lls completely,
the meter will start counting down. Remove the control solution
sample from the test strip when the countdown begins.
Note: To avoid contaminating the entire vial of control solution, do
not directly apply control solution onto a strip.
Step 4: Read and compare the results.
After the meter counts down to "1"; the control solution test result
will appear on the meter display.