Beurer GL 44 LEAN mg/dL

Product's Documents

Below are documents related to this product, you can read online or download:
Installation Instruction Other Documents
  • Instruction for use GL 44 lean mg/dL English EN - (English) Download
Specification
  • Data sheet GL 44 lean mg/dL English EN - (English) Download
GL 44 LEAN photo

Quick start guide GL 44 lean mg/dL English EN

This is the main product document for model GL 44 LEAN. Additionally, the document applies to other Beurer models: 4211125463083, 46308

The file format is pdf, 8 pages, you can download this manual here .

background
GL 44 lean
ENshort instruction
Lancet needles / Lanzetten / lancettes / lancetas / lancette:
SteriLance Medical (Suzhou) Inc.,No.168 PuTuoShan Road
New District, 215153 Suzhou, Jiangsu
PEOPLE’S REPUBLIC OF CHINA
Emergo Europe B.V., Westervoortsedijk 60
6827 AT Arnhem, The Netherlands
GL 44 lean / LD 04
Beurer GmbH • Söflinger Straße 218 • 89077 Ulm, Germany • www.beurer.com
www.beurer-gesundheitsratgeber.com • www.beurer-healthguide.com
background
2
Warnings and safety notes
Before first use
Always read the full instructions for use before taking your first blood glucose measure-
ment. Check the measuring unit in accordance with the instructions for use.
Risk of infection
The monitor and accessories may come into contact with human blood and are therefore
a possible source of infections.
Never use a monitor, lancing device or accessories together with other people.
When using the measurement system on dierent people, please observe the general
hygiene regulations.
Use a new, sterile lancet needle for taking each blood sample.
Prior to taking the blood sample, wash your hands with warm water and soap, and
then dry them.
background
3
Preparing the monitor and lancing device
Inserting the test strip
Insert the test strip into the monitor with the contacts first.
The start screen is shown for a short moment. Then wait briefly until the
flashing hand and blood droplet symbol can be seen on the display.
Inserting the lancet needle
Remove cap from the lancing device. Insert lancet needle in the lancing device.
Hold the shaft of the lancet needle and remove protective disc by turning.
Retain the protective disc. Reapply the cap to the lancing device. Turn the
moving upper part of the cap until the desired penetration depth is dis
-
played. Pull the tensioning device back until it engages.
background
4
Taking a blood sample
Pricking your finger
Firmly position the lancing device slightly to the side of the centre of the
fingertip. Press the trigger.
The blood sample taken must:
be in droplet from and must not be spread
be at least 1.4 mm (original size ).
5
background
5
Applying blood to the test strip
Hold the blood input gap on the test strip against the drop of blood until the
gap is completely filled and the display of the monitor starts counting back
-
wards. Wait for approx. 5 seconds until the blood glucose value measured is
shown in the display.
Normal value: 2 hours after a meal: 70 140 mg/dL (3.9 7.8 mmol/L)
On an empty stomach: 70 100 mg/dL (3.9 5.6 mmol/L)
Disposing of the lancet needle
Remove cap. Place safety shield on a flat surface and push the tip of the
needle into it. Use the sliding switch on the side. This causes the lancet to
drop out of the holder. Properly dispose of test strip and lancet. Reapply
cap.
background
6
Evaluating critical measured values
Very low blood glucose level
< 20 mg/dL (< 1.1 mmol/L)
Seek medical attention im-
mediately.
Low blood glucose level
< 70 mg/dL (< 3.9 mmol/L)
Have a snack.
Follow your GP's instructions.
High blood glucose level
On an empty stomach > 100 mg/dL
(> 5.6 mmol/L)
2 hours after a meal > 140 mg/dL
(> 7.8 mmol/L)
Seek medical attention.
background
7
High blood glucose level, possibly
ketones
> 240 mg/dL (> 13.3 mmol/L)
Ketone test recommended.
Consult your GP.
Very high blood glucose level
> 630 mg/dL (> 35.0 mmol/L)
Take another measurement
using a new test strip. If the
result is identical, seek medi
-
cal attention immediately.
background
8
Information on the display
GL44lean_2024-07-03_03_QSG_BEU_EN Subject to errors and changes
LP
Batteries empty. Replace all batteries.
Ht or Lt
Temperature is outside the
permitted range.
Repeat the test with a new strip at room
temperature +20 °C to +26 °C.
Err
Used or contaminated test
strip.
Insert a new test strip that has not expired
and repeat the measurement.
Err001/005
System error.
Remove batteries. Insert batteries again.
Should the problem persist, contact Cus-
tomer Services.
Err002
Test strip not filled suciently.
Repeat the measurement using a new test
strip.

Specifications