FLUKE 717 30G Pressure Calibrator, 12 to 30 PSI

User Manual for the Fluke 717 30G Pressure Calibrator, 12 to 30 PSI - Page 2

For 717 30G.

PDF File Manual, 16 pages, Read Online | Download pdf file

717 30G photo
Loading ...
English
Input Units
The Calibrator measures and displays pressure sensor inputs in the units listed
below:
Psi
inH2O@4°C
inH2O@20°C
cmH2O@4°C
cmH2O@20°C
bar
mbar
kPa
inHg@0°C
mmHg
kg/cm
2
If inappropriate units are selected, the output of Fluke 700P pressure modules can
be too low to be displayed or can cause the Calibrator to display OL (overload).
Refer to Table 1 for pressure unit and range compatibility.
Table 1. Pressure Unit and Range Compatibility
Units Range Units Range
psi All kPa All
inH
2
O Through 3000 psi inHg All
cmH
2
O Through 1000 psi mmHg Through 3000 psi
bar 15 psi and above kg/cm
2
15 psi and above
mbar Through 1000 psi
Symbols
Symbol Meanin g
+
ON / OFF button
J
Earth ground
W
Caution: Important information. See instruction sheet
X
Hazardous voltage, risk of electric shock
f
Pressure
T
Double insulated
M Battery
$
Canadian Standards Association
P
Conforms to European Union requirements
F
Direct current