Maintenance
XW Warning
To avoid electric shock, personal injury, or damage to the Calibrator:
Do not service this product other than as described in this
Instruction Sheet unless you are a qualified technician and have
the required equipment and service information.
Remove any input signals prior to removing test leads and opening
case.
When servicing the Calibrator, use only specified replacement
parts.
Do not allow water to get in the case.
For maintenance procedures not described in this Instruction Sheet, contact a
Fluke Service Center.
In Case of Difficulty
Check the battery, test leads, and pressure tubing. Replace as necessary.
Review this Instruction Sheet to make sure you are using the Calibrator
correctly.
If the Calibrator needs repair, and the Calibrator is under warranty, see the
warranty statement for terms. If the warranty has lapsed, the Calibrator will be
repaired and returned for a fixed fee.
Cleaning
Periodically wipe the case with a damp cloth and detergent; do not use abrasives
or solvents.
Calibration
Calibrate your Calibrator yearly to ensure that it performs to specification. A
Calibration Manual (Fluke PN 686540) is available.
Replacing the Battery
XW Warning
To avoid false readings, which could lead to electric shock or injury,
replace the battery as soon as M (low battery indicator) appears on
the display.
Use only a single 9 V battery, properly installed, to power the
Calibrator.
The Calibrator uses a single 9 V, alkaline battery (ANSI/NEDA 1604A or IEC
6LR61).
To replace the battery (see Figure 5):
1. Turn the Calibrator OFF and remove the test leads from the terminals.
2. Remove Calibrator from its holster.
3. Remove the battery door on the back of the Calibrator as shown.
4. Lift the battery from the battery receptacle.
5. Attach the replacement battery to the leads, place the battery back in the
battery receptacle.
6. Secure the battery door.
7. Return the Calibrator to its holster.