probes be inserted to an adequately depth into the material being mea
-
sured. Measuring temperature using a surface sensor can be especially
difficult as the sensor is directly exposed to ambient.
Error Message at Power Up
The 1502A reports an error during the power up self-test.
On power up the 1502A performs a self-test of several of its key components.
A failure of a component will cause an error message to be displayed such as
“Err 4”. The possible error messages and their meanings are as follows:
Err 1
Static RAM failure.
Err 2
Nonvolatile RAM failure.
Err 3
Internal data structure error.
Err 4
ADC initialization failure.
Err 5
ADC operation error.
Generally, each of these conditions require a qualified factory technician to re-
place a faulty component. Contact the factory for assistance. One possible ex-
ception might be if a large static discharge nearby disturbs the circuits. Cycling
the power off and back on again may allow the 1502A to resume normal oper-
ation. Another might be if the AC source voltage is incorrect, e.g. using 115 V
when the 1502A is configured for 230 V. Check the source voltage and the
1502A’s configuration and make sure they agree.
38 Manual Rev. 952801 Hart Scientific
10 Troubleshooting
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com