7.3.1.1 Reading Temperature
The most recent temperature measurement can be read using the following
command:
T<EOS> reads the most recent measurement
The syntax of the response is as follows:
t:_nnnn.nnn_u
or
t:_nnnn.nnn_u_hh:mm:ss
The _’s represent space characters. The n’s represent the digits of the mea-
surement value. If fewer digits are needed the leading positions are filled with
space characters. The u represents the unit which is either ‘C’, ‘F’, ‘K’, or ‘O’
(for ohms). The time stamp appears if this option is enabled (see Section
7.3.1.4 below). The time appears in 24-hour format with two digits each for
hours, minutes, and seconds.
The following SCPI compatible command can also be used to return the most
recent measurement but without the label and unit.
FETC?<EOS> or
FETCH?<EOS> returns the value of the most recent measurement
28 Manual Rev. 952801 Hart Scientific
7 Digital Communications Interface
Command Description
Communication parameter commands
DU[=F/H] read [or set] serial sample duplex mode
LF[=ON/OF] read [or set] serial linefeed
Calibration Commands
*PA=<password> disable password lockout of calibration commands
*LO=[=CA|AL] read [or set] menu lockout
*C0[=<value>]
read [or set] the 0Ω calibration parameter
*C1[=<value>]
read [or set] the 100Ω calibration parameter
*C4[=<value>]
read [or set] the 400Ω calibration parameter
*SN[=<value>] read [or set] the instrument serial number
Miscellaneous Commands
*VER read model number and firmware version number
IDN? read manufacturer, model number, serial number, and firmware
version number (SCPI compatible)
H read a list of commands
Command List Continued