Error Messages and Troubleshooting
Error Messages 7
7-15
Table 7-1. Error Messages (cont.)
Error Code Error Message Causes and Solutions
Error 651
USB memory not
ready
Cause: USB memory is not ready for data record.
Solution: You may have configured the record destination to
USB, but the memory is not ready when recording is started.
Error 652
Data file limit
exceeded
Cause: Too many data files were created for a scan.
Solution: Start a new scan.
Error 653
Error creating record
file
Cause: Failed to create file for data record.
Solution: Make sure that the destination memory is not full.
Error 654
Error writing record
file
Cause: Failed to write file for data record.
Solution: Make sure that the destination memory is not full.
Error 655
Error resuming data
record
Cause: Failed to resume the data record.
Solution: Make sure that the destination memory is not full.
Error 656
Data record limit
exceeded
Cause: Too many data records in internal memory
(Max.1000).
Solution: Remove some data records.
Error 657 Incorrect timestamp
Cause: RTC clock is reset.
Solution: Set the instrument’s date time setting.
Error 658
Record state
damaged
Cause: The battery could have failed.
Solution: Replace the battery.
Error 700 NV memory invalid
Cause: The flash memory contains invalid Instrument setup,
LAN setup, MAC setup or calibrate constants.
Solution: Default settings will be loaded, if the MAC setup or
calibrate constants show failed, Contact Fluke. See Chapter
1.
Error 701
NV invalid so default
loaded
Cause: Default settings have been loaded.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 702
Error reading
instrument setup
Cause: Internal flash is damaged.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 703
Error writing
instrument setup
Cause: Internal flash is damaged.
Solution: Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).