Error Messages and Troubleshooting
Error Messages 7
7-13
Table 7-1. Error Messages (cont.)
Error Code Error Message Causes and Solutions
Error 431 Invalid setpoint
Cause: Setpoint that is specified in the command is not in
range 1 to 20. This could be because of a wrong setpoint in
the TRIGger:AUTO:POINts, TRIGger:AUTO:SPOint,
TRIGger:AUTO:TOLerance, TRIGger:AUTO:STABility,
TRIGger:AUTO:DWELl commands
Solution: Check the parameter value and try again.
Error 432
Input module not
installed
Cause: The Input Module is not installed.
Solution: Install the Input Module and try again.
Error 433
Recording is
automatic
Cause: Recorder is set to OFF while automatic recording is in
progress.
Solution: Stop the scan or use the ABORt command.
Error 434
Cannot record while
idle
Cause: Recorder is started, but a scan has not been started
yet.
Solution: Use command INITiate to start scan.
Error 435
Only fixed CJC
allowed on channel
Cause: Ch001 is being configured to thermocouple with
internal or external CJC, that is not supported on this channel.
Solution: Set up the input on a different channel.
Error 436
Unavailable
operation
Cause: The operation is not supported on specified channel.
Solution: Check the parameter value and try again.
Error 439
DIO currently
committed to scan
Cause: DIO cannot be used as output while it is included in
scan list.
Solution: Configure the setpoints.
Error 437 Setpoints not set
Cause: User tried to start an Automated Test without
setpoints.
Solution: Use command *RST to reset to factory default, or
set the DIO channel to OFF.
Error 440
Not able to achieve
requested resolution
Cause: The Instrument cannot achieve the requested
measurement resolution. Invalid resolution in the CONFigure
or MEASure? Command could have been set.
Solution: Check the parameter value and try again.
Error 441
Cannot monitor math
channel
Cause: The user tried to monitor a math channel. Math
channels cannot be monitored.
Solution: See the math channel data in the scan data screen.
Error 501
Invalid keyword or
choice
Cause: Invalid keyword is found in the command.
Solution: Check the parameter value and try again.