Micsig TO2004 oscilloscope

Micsig TO004 series user manual - Page 64

For TO2004.

PDF File Manual, 355 pages, Read Online | Download pdf file

TO2004 photo
Loading ...
Loading ...
Loading ...
Chapter 3 Horizontal System
61
1) Connect sine wave signals to CH1 and connect sine wave signals of the same frequency and different phases to
CH2.
2) Press “Autoset button, tap “Displayin the main menu, then select “XYin “Time Base”.
3) Drag signals so that they are centered on the display screen. Adjust the vertical sensitivity of CH1 and CH2,
and extend signals for viewing.
The phase difference (θ) can be calculated using the following formula (assuming that the amplitudes of the
two channels are the same):

