time the record length is 220/110K In the case of, the maximum number of segments allowed is 1791, so the
number of segments for segmented memory will be automatically set to 1791, and the user can also modify the
number of segments required between 1-1791.
10.5.2 Acquire Waveform
There are two collection methods, conventional collection and single collection.
Routine acquisition, including automatic and normal, each acquired waveform is displayed on the screen in time,
occupying one of the segments, and after all segments are used up, it is stored in a loop.
Single acquisition, no display during the acquisition process, display only after the acquisition is stopped; one
acquisition, fill all segments, and stop the acquisition after the acquisition (all segments are filled); during the
acquisition process, you can press the "stop" button, Stop collection.
Note: In segmented memory, it is recommended to set the trigger mode to normal to store more effective signals.
The current segment/total allowable segment number during the acquisition process is displayed in the upper left
corner of the waveform area.