Ti9, Ti10, Ti25, TiRx, TiR and TiR1
Users Manual
14
Making Accurate Temperature Measurements
All objects radiate infrared energy. The amount of energy radiated is based on
two primary factors: the surface temperature of the object and the emissivity of
the object's surface. The Imager detects the infrared energy from an object and
uses this information to estimate the temperature of the object. Most of the
objects measured such as painted metal, wood, water, skin, and cloth are very
efficient at radiating energy and it is easy to get very accurate measurements.
For surfaces that are efficient at radiating energy (high emissivity), the
emissivity factor is estimated to be 95% (or 0.95). This estimate works for
most purposes. This simplification does not work however, on shiny surfaces
or unpainted metals. These materials are not efficient at radiating energy and
are classified as having low emissivity. To accurately measure materials with a
low emissivity, an emissivity correction is necessary. The easiest correction
method is to set the Imager to the proper emissivity value so the Imager
automatically calculates the correct surface temperature. If the Imager uses a
fixed emissivity value (meaning that it is set to one value and the user cannot
adjust it), then the Imager’s measurement must be multiplied by a value found
in a look up table to get a more accurate estimate of the actual temperature.
The TiR1 and Ti25 have the ability to set emissivity by directly entering a
value or using a table of built-in values. The TiRx, TiR, Ti9, and Ti10 have
emissivity that is permanently fixed at 0.95, which works well for most
surfaces but can provide very inaccurate reading if used directly on a shiny
metal surface.
A significant amount of information is available on emissivity. Further study
of this topic is recommended to get the most accurate temperature
measurements using the Imager.
Setting Emissivity (Ti25 and TiR1 Only)
Setting your Imager with correct emissivity values is critical to making
accurate temperature measurements. To set the emissivity value:
1. Press a until Emissivity appears over c.
2. Press the softkey labeled Emissivity.
At this point, emissivity can be set directly as a value or selected from a list of
emissivity values for some common materials. To select from a list of common
materials:
1. Press the softkey labeled Table.