NORMA 4000/5000
Operators Manual
9-8
Adjust the settings:
Column Settings Description
Ref
FIX
U1, M1, P
M1
Fixed DC voltage, or selection from available average
measured values
Gain 1… Transducer ratio or fixed value (-10.3 V to +10.3 V)
Unit V/A, V/V, V/Ohm, V/Hz
(depending if Ref is not
selected)
that is 10 mV/V, for example, 10 mV at the output
corresponds to 1 V of the measured value
Zero 1... Set zero/offset
Unit A, W, V, Hz, Ohm Unit for zero, depending on selected Ref
2. Move the cursor to a field in line A1 and press Enter.
A list of possible options is displayed.
3. Select a value and press Enter to confirm.
The value is now shown in the display field.
4. Configure analog outputs A2 to A4 accordingly.
Measuring with the Process Interface
Torque, rotational speed and mechanical power are measured in real-time and averaged.
They are combined with the measured electrical values so that slip and mechanical
efficiency can be calculated.
The device must be equipped with an Analog Interface.
Menu item PI must be visible in the menu bar.
If the Power Analyzer is equipped with a process interface, menu item PI is shown
automatically in the menu.
View Measured Electric Values
1. Press the  measuring key (numerical display).
The measured values of channel 1 are shown.