SIGLENT
SSA3000X User Manual 48
2.3.2.2 M -> CF Step
Set the center frequency step of the analyzer to the frequency of the current marker.
If Normal marker is selected, the center frequency step will be set to the frequency of the
current marker.
If Delta or Delta Pair marker is selected, the center frequency step will be set to the frequency
of the Delta Marker.
The function is invalid in Zero span mode.
2.3.2.3 M -> Start Freq
Set the start frequency of the analyzer to the frequency of the current marker.
If Normal marker is selected, the start frequency will be set to the frequency of the current
marker.
If Delta or Delta Pair marker is selected, the start frequency will be set to the frequency of the
Delta Marker.
The function is invalid in Zero span mode.
2.3.2.4 M -> Stop Freq
Set the stop frequency of the analyzer to the frequency of the current marker.
If Normal marker is selected, the stop frequency will be set to the frequency of the current
marker.
If Delta or Delta Pair marker is selected, the stop frequency will be set to the frequency of the
Delta Marker.
The function is invalid in Zero span mode.
2.3.2.5 M ->Ref Level
Set the reference level of the analyzer to the amplitude of the current marker.
If Normal marker is selected, the reference level will be set to the amplitude of the current
marker.
If Delta or Delta Pair marker is selected, the reference level will be set to the amplitude of the
Delta Marker.
2.3.2.6 ΔM->Span
Set the span of the analyzer to the frequency difference between the two markers in Delta, or
Delta Pair marker type.
If Normal marker is selected, this function is invalid.
The function is invalid in Zero span mode.