SIGLENT
SSA3000X User Manual 36
Decreasing the sweep time would speed the measurement. However, an error may be caused
if the specified sweep time is less than the minimum sweep time in Auto coupling; at this point,
UNCAL‖ is shown in the status bar on the screen
Table 2-16 Sweep Time
Parameter
Explanation
Default
N/A
Range
900us ~ 3ksQuasi Peak900us
~ 30ks
Unit
ks, s, ms, us
Knob Step
Sweep time/100, min =1 ms
Direction Key Step
in 1,3 sequence
2.2.4.2 Sweep Rule
The analyzer provides two sweep time rules to meet the different sweep time requirements:
Speed: Activates the default fast sweep time rule.
Accuracy: Activates the normal sweep time rule to ensure the measurement accuracy. Speed
sweep time rule provides a fast measurement function that decreases the sweep time. Using Fast
Sweep will decrease the measurement accuracy.
2.2.4.3 Sweep
Set sweep mode in single or continuous, the default is continuous. The corresponding icon of the
sweep will be displayed in the status bar at the left of the screen.
1. Single
Set the sweep mode to Single‖. The number on the parameter icon denotes the current sweep
number.
2. Numbers
Set the number of sweeps for a single sweep. In single sweep mode, the system executes the
specified number of sweeps and the number shown on the icon in the status bar at the left of the
screen varies with the process of the sweep.
3. Continue
Set the sweep mode to ―Continue‖. The character Cont on the parameter icon denotes the
analyzer is sweeping continuously.
If the instrument is in single sweep mode and no measurement function is enabled, press this
key and the system will enter continuous sweep mode and sweep continuously if the trigger
conditions are satisfied.
If the instrument is in single sweep mode and a measurement function is on, press this key
and the system will enter continuous sweep mode and measure continuously if the trigger