Specifications
Electrical Specifications
Current Measurement
With Jaw
Ranges ................................ 0-20.99 mA; 21-100 mA
Resolution............................ 0.01 mA; 0.1 mA
Accuracy.............................. 0.2 % + 5 counts;1 % + 5 counts
In Circuit
Range ..................................0-24 mA
Resolution............................ 0.01 mA
Accuracy.............................. 0.2 % + 2 counts
Current Source
Range ..................................0-24 mA
Resolution............................ 0.01 mA
Accuracy.............................. 0.2 % + 2 counts
mA Drive.............................. 24 mA into 1000
Current Simulate
Range ..................................0-24 mA
Resolution............................ 0.01 mA
Accuracy.............................. 0.2 % + 2 counts
Maximum Voltage................ 50 V
DC Voltage Measurement (773)
Range ..................................0-30 V
Resolution............................ 0.01 V
Accuracy.............................. 0.2 % + 2 counts
DC voltage source (773)
Range ..................................0-10 V
Resolution............................ 0.01 V
Accuracy.............................. 0.2 % + 2 counts
mA Drive.............................. 2 mA max all conditions
mA IN/OUT (773)
Sourcing range .................... 0-24 mA
Sourcing resolution.............. 0.01 mA
Sourcing accuracy ............... 0.2 % + 2 counts
Measurement range.............0-24 mA
Measurement resolution ...... 0.01 mA
Measurement accuracy ....... 1 % FS
Scaled mA current output to mA current input from the Jaw (773)
Range ..................................0-24 mA
Resolution............................ 0.01 mA
Accuracy.............................. 1 % FS
Response speed ........................ 2 times per second
DC Loop Power.......................... 24 V
Influence of Earth’s Field............ <0.20 mA
Batteries.....................................4 1.5 V, Alkaline , IEC LR6
Working hours............................ 12 hours @ 12 mA sourced into 500
Mechanical Specifications
Size (H X W X L) .......................43.7 mm x 70 mm x 246.2 mm
Weight........................................410 g