5-8
Using Color Alarms
Note
This feature is not available on the TiR3 or Ti50.
The color alarm function allows you to highlight target object areas of thermal
interest by selectively fusing the visible light image with portions of the
infrared image. For example, you can specify a temperature range. Then, any
target object temperatures either inside or outside of the range, depending on
what you set, are highlighted in the infrared color that corresponds to that
temperature (based on the color palette settings).
You can choose to highlight temperatures either inside or outside of the range
specified. You can also choose to highlight only temperatures above or below a
threshold setting.
Color alarm thresholds and range limits can be adjusted using the display
screen menu.
Note
Color alarms work in both full screen and picture-in-picture view.
With the onscreen alarm disabled, the entire infrared image is
blended with the visible light image according to the IR-Fusion
®
blend setting and palette selection. The IR-Fusion
®
blend setting
should include some IR (i.e., not full VL) so that the color alarm is
visible on the display screen, provided objects in the scene meet the
color alarm criteria.
Using Menu to Adjust Color Alarm Ranges
To enable/disable onscreen alarm:
1. Tap F.
2. Use the mouse controller to position the pointer over Image Settings on
the popup menu and tap E.
3. Position the pointer over the Color Alarms tab and tap E. You may
need to position the pointer over the right or left arrow and tap E to
scroll to the Color Alarms tab.