ScopeMeter® Test Tool 190 Series II
Users Manual
122
AC+DC Voltage (True RMS)
Ranges ..... 500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V
Full Scale Reading ........................................ 5000 counts
Accuracy
DC to 60 Hz ..................................... ±(1 % +10 counts)
60 Hz to 1 kHz ...............................±(2.5 % +15 counts)
For higher frequencies the frequency roll off of the Meter
input starts affecting accuracy.
All accuracies are valid if the waveform amplitude is
larger than 5 % of full scale.
Amperes (AMP)
With Optional Current Probe or Current Shunt
Ranges ............................. same as VDC, VAC, VAC+DC
Probe Sensitivity ................. 100 µV/A, 1 mV/A, 10 mV/A,
100 mV/A, 1 V/A, 10 V/A, and 100 V/A
Accuracy ........................... same as VDC, VAC, VAC+DC
(add current probe or current shunt accuracy)
Recorder
TrendPlot (Meter or Scope)
Chart recorder that plots a graph of min and max values of
Meter or Scope measurements over time.
Measurement Speed ......................... >5 measurements/s
Time/Div ............................................ 5 s/div to 30 min/div
Record Size (min, max, average) ................. 19200 points
Recorded Time Span ........................ 64 min to 546 hours
.........................................................................................
Time Reference ..................... time from start, time of day
Scope Record
Records scope waveforms in deep memory while
displaying the waveform in Roll mode.
Source ...................................................... Input A, B, C, D
Max. Sample Speed (4 ms/div to 1 min/div) ....... 125 MS/s
Glitch capture (4 ms/div to 2 min/div) ......................... 8 ns
Time/Div in normal mode ................. 4 ms/div to 2 min/div
Record Size ............................... 30k points per waveform
Recorded Time Span ............................. 4.8 s to 40 hours
1.888.610.7664 sales@GlobalTestSupply.com