NORMA 4000/5000
Operators Manual
8-4
Measure Voltage, Current and Power
Measured Values for Individual Channels
Note
The description that follows is for a W3 configuration. The W2
configuration is basically the same with some differences. In W2, some
values are invalid and are suppressed. The phase voltage does not exist and
is replaced by the phase-to-phase voltage, reactive power, and apparent
power and are available as totals only.
View the Values of One Channel
After switching on the Power Analyzer, the display shows the numerical values measured
in channel 1.
esn030.gif
Display Description
U
1 rms
rms voltage value
I
1 rms
rms current value
P
1
Real power
S
1
Apparent power
Q
1
Reactive power
λ
1
Power factor lambda (capacity or inductance)
Press measuring keys 1 through n to view the values of the respective channels.
View Detailed Values of One Channel
You have the option to view detailed data regarding the measured values of a channel.
1. Press measuring keys 1...n to view the measured values of the respective channel.
2. Press function key Detail.
Details regarding the voltage values for channel 2 are shown in the following example.