Accuracy Tests
9
Accuracy Tests
Accuracy specifications are valid for 1 year after calibration adjustment when measured
at an operation temperature of 18 °C to 28 °C. Allow the Meter to stabilize at room
temperature prior to performing the accuracy tests.
The following tables list the required performance test points for verifying Meter
accuracy. Zero the Meter prior to completing each measurement point.
mA DC Clamp Measure Accuracy Tests
UUT Meter Reading Limit
Step Calibrator Output
Low High
1 4.00 mA 3.94 mA 4.06 mA
2 -4.00 mA -4.06 mA -3.94 mA
3 12.00 mA 11.03 mA 12.07 mA
4 -12.00 mA -12.07 mA -11.03 mA
5 20.00 mA 19.01 mA 20.09 mA
6 -20.00 mA -20.09 mA -10.01 mA
7 100.0 mA 98.5 mA 101.5 mA
8 -100.0 mA -101.5 mA -98.5 mA
mA DC Measure Accuracy Tests
UUT Meter reading limit
Step Calibrator Output
Low High
1 0.00 mA -0.02 mA 0.02 mA
2 4.00 mA 3.97 mA 4.03 mA
3 -4.00 mA -4.03 mA -3.97 mA
4 8.00 mA 7.96 mA 8.04 mA
5 -8.00 mA -8.04 mA -7.96 mA
6 12.00 mA 11.96 mA 12.04 mA
7 -12.00 mA -12.04 mA -11.96 mA
8 20.00 mA 19.94 mA 20.06 mA
9 -20.00 mA -20.06 mA -19.94 mA
10 24.00 mA 23.93 mA 24.07 mA
11 -24.00 mA -24.07 mA -23.93 mA