FLUKE N4K3PP42B N4K 3PP42B Norma 4000 3-Phase Power Analyzer with 3 x PP42 Modules with Binding Posts

User Manual for the Fluke N4K 3PP42B Norma 4000 3-Phase Power Analyzer with 3 x PP42 Modules with Binding Posts - Page 127

For N4K3PP42B.

PDF File Manual, 161 pages, Read Online | Download pdf file

N4K3PP42B photo
Loading ...
Loading ...
Loading ...
NORMA Process Interface (Optional)
Configuring the Process Interface 9
9-7
1. Move the cursor to the field in line Drv1 and press Enter.
A list of possible options is displayed.
2. Select a value and press Enter to confirm.
The value is now shown in the display field.
3. Press function key SAVE to save this configuration.
Configure Other Motors
1. Press the Next... function key.
The settings for motor 2 are displayed.
2. Adjust settings for motors 2 to 4, following the above instructions for motor 1.
3. Press SAVE to save the configurations for the motors.
Configure Analog Output
The 4 analog outputs (A1...A4) can be used to output the values measured, calculated, or
averaged, or to transfer them to an external device for further processing. By default, the
analog outputs are configured as voltage output for ±10 V. In order to output higher
voltages, you must enter the relevant transducer ratio, for example, 10 mV/V for a
measured voltage of 220 V and an output of 2.2 V.
1. Press function key A-Out.
The Analog Output Setup menu is displayed.
esn065.gif
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com
Loading ...
Loading ...
Loading ...