Connection to Circuits
3-Phase Measurement (W3) 5
5-13
3-Phase Measurement (W3)
Direct Connection
Figure 5-8 illustrates a direct connection for a 3-phase measurement. Ensure that there is
no overload at the current input of the Power Analyzer. If there is a potential risk of
overload at the current input, incorporate a shunt or transducer into the circuit. If
necessary, install appropriate fuses.
To select the 3-phase (W2) method, go to General Setup and select W3. For more
information, see Chapter 7, “Configuration.”
XW Warning
To avoid possible electric shock or personal injury:
Risk of injury when touching connections, internal circuits
and measuring devices that are not connected to earth
ground.
Always adhere to the instructions regarding the sequence of
connection (see chapter 5, Connecting Sequence).
HI
VOLTAGE
1000 V max
LO
PROBE
EXT.SHUNT
10 V max
123
HI
CURRENT
10 A max
LO
ALL INPUTS MAX 1000V CATII TO
PROBE
EXT.SHUNT
10 V max
HI
CURRENT
10 A max
LO
HI
VOLTAGE
1000 V max
LO
L1
L2
L3
N
Max. 10 m
Load
I
3
I
2
I
1
V
3
V
2
V
1
esn015.eps
Figure 5-8. 3-Phase Measurement-Direct Connection