NORMA Process Interface (Optional)
Configuring the Process Interface 9
9-5
1. Move the cursor to menu item PI and press Enter.
Menu Motor / Generator Setup is displayed, showing the settings for motor 1
(M1) as shown below.
esn064.gif
Adjust the settings as follows:
Line Description
M1 Configure torque measurement (input, slope and zero) for each motor
n1 Configure speed measurement (speed sensor)
Drv1 Set type (Type), pole pairs (PPairs) and reference power (Pref)
2. If a configuration that suits the measuring layout is already saved, press function
key LOAD, select the configuration.
3. Press Enter to confirm.
4. Adjust configuration as described in the section that follows.
Select Motor
To configure the system for motor 1, go to the “Configure Torque Sensor”
section.
To configure another motor, press Next ... until the respective motor code (M2,
M3 or M4) is displayed.